Description
Agilent 1100 Series HPLC System
This item comes with a 30-day Repair or Replace warranty.  Â
This HPLC system is in very good condition. All operational testing was completed successfully, and each unit functions as intended. We used in-house software for testing. Please Note: If you need computer and software, please contact us for information prior to ordering, otherwise will come only with what is shown.
G1313A – Autosampler Â
- Injection Range: 0.1 µL to 100 µL, with 0.1 µL increments Â
- Replicate Injections: 1 to 99 per vial Â
- Sample Capacity: 100 x 2 mL vials per tray Â
- Minimum Sample Volume: 1 µL from a 5 µL sample in a 100 µL micro vial Â
- Injection Cycle Time: Typically, 50 seconds Â
- Carryover: Typically, < 0.01% to < 0.05% Â
- Sample Viscosity Range: 0.2 cp to 50 cp Â
- Physical Dimensions: 200 mm (height) x 345 mm (width) x 435 mm (depth)Â Â
- Weight: 14.2 kg (31.3Â lbs)Â Â
- Power Requirements: 100-120 or 220-240 VAC, ± 10%, 50 or 60 HzÂ
G1316A – Thermostatted Column Compartment (TCC)
- Temperature Range: 10°C below ambient to 80°C Â
- Temperature Stability: ± 0.15°C Â
- Temperature Accuracy: ± 0.8°C Â
- Column Capacity: Three 30 cm columns Â
- Warm-up/Cool Down Time: 5 minutes from ambient to 40°C  Â
G1315B – Diode Array Detector (DAD)Â
- Detection Type: 1024-element photodiode array Â
- Light Source: Deuterium and tungsten lamps Â
- Wavelength Range: 190 – 950 nm Â
- Short Term Noise: ± 1 x 10^-5 AU at 254 and 750 nm Â
- Drift: 2 x 10^-3 AU/hr at 254 nm Â
- Linear Absorbance Range: > 2 AU (upper limit)Â Â
- Wavelength Accuracy: ± 1 nm Â
- Slit Width: Programmable slit widths of 1, 2, 4, 8, and 16 nm Â
- Flow Cells: Various options including standard, semi-micro, micro, high pressure, and nano flow cells Â
- Control and Data Evaluation: Agilent ChemStation for LCÂ Â
- Analog Outputs: Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, two outputs Â
- Communications: Controller-area network (CAN), GPIB, RS-232C, APG Remote, LAN optional Â
- Safety and Maintenance: Extensive diagnostics, error detection, leak detection, and safe leak handling Â
- Power Requirements: 100 – 240 VAC, 50 or 60 Hz, 300 VA / 125 W / 427 BTUÂ
G1312A – Binary Pump Â
- Weight: 34 lbs. Â
- Dimensions: 7 x 13.5 x 17 in.Â
- Electrical: 100 – 240 VAC, ± 10% ~ 50/60Hz ~ 220VA ~ 74W ~ 253BTU Â
- Ambient Operating Temp: 4–55 °C (41–131 °F) Â
- Ambient Non-Operating Temp: -40–70 °C (-4–158 °F) Â
- Maximum Operating Pressure: 600 bar (8700 psi)Â Â
- Flow Rate Range: 0.1 – 5 mL/min Â
- Flow Precision: < 0.07% RSD or < 0.02 min SDÂ Â
- Flow Accuracy: ± 1% or 10 µL/minÂ
G1379A – Micro Vacuum Degasser Â
- Channels: 4-channel vacuum container Â
- Membranes: 4 micro-structured membranes Â
- Vacuum Pump: Integrated Â
- Internal Volume: 480 µL per channel Â
- Maximum Flow Rate: 10 mL/min per channel Â
- Operating Temperature: 4–55 °C (41–131 °F) Â
- Power Consumption: 30 VAÂ