Amersham Pharmacia Biotech AKTA Explorer System with Computer and Software

$11,000.00

1 in stock

Description

 

The attached
photos represent an AKTA Purifier System with Computer and Software
being sold as is with the understanding that we sell quality products
and equipment that operate as intended unless otherwise stated. Please
review my feedback to see that I have a solid record.
 
Comes
with an Amersham ATKA Box-900, a pH/C-900 Monitor, a UV-900 Monitor, a
P-900 Pump, a M-925 Mixer, seven (7) INV-907 Valves, a P-950 Sample Pump, a
CU-950 Control Unit, with Unicorn Software already installed on the laptop.

pH/C-900 Monitor:

 

  • Accuracy:
    The greater of ± 2% of full scale or ± 10 µS/cm in the range 1
    µS/cm-300 mS/cm ± 0.1 pH unit, temperature compensated Max. 0.1 pH units
    within 0–10 ml/min (pH measurement)

     

  • Operating pressure max.: 50 bar (5 MPa) (725 psi)

     

  • Conductivity: 1 µS/cm-999.9 mS/cm

     

  • Operating Temperature: 4°C-40°C

     

  • Relative humidity: 20%-95% (non condensing)

     

  • Baseline noise: ± 0.5% of full scale calibrated range

     

  • Drift: Max. 0.02 pH/h (measured at pH 4.0) ±0.5 °C per 10 h 0.1 pH unit deviation/10 h

     

  • Cell Constant: 50 ± 20 cm-1

     

  • Response Time: Max. 10 s (0–95% of step) (pH measurements) Max. 3 s (0–95% of step) (Conductivity measurement)

     

  • Voltage: 100-120/220-240 VAC

     

  • Protection class: IP 20 (Housing) IP 44 (Flow cell)

     

  • Material:
    Glass Perfluororubber (FFKM) Polyetheretherketone (PEEK)
    Polytetrafluoroethylene (PTFE) Titanium alloy (Ti alloy)
    Chlorotrifluoroethylene (CTFE)

     

 

UV-900 Monitor:

 

  • Accuracy: ±2 nm
  • Operating pressure max.: 2 MPa
  • Operating Temperature: 4°C-40°C
  • Relative humidity: 20%-95% (non condensing)
  • Linearity: < 2% deviation up to 2 AU at 260 nm with Uracil at pH 2
  • Baseline noise: < 60 μAU at 230 nm
  • Drift: < 2×10-4 AU/h at 254 nm
  • Voltage: 100-120/220-240 VAC
  • Protection class: IP 20
  • Material: Polyetheretherketone (PEEK) Polytetrafluoroethylene (PTFE) Titanium alloy (Ti alloy) Glass 

 

P-900 Pump:

 

  • Accuracy: ±2% or 20 μl/min whichever is greater ( 0.2–10.0 MPa)

     

  • Analog Output: Pressure, 0 to 1V

     

  • Chemical
    stability: The wetted parts are resistant to organic solvents and salt
    buffers commonly used in chromatography of biomolecules, except 100%
    Ethylacetate, 100% Hexane, and 100 % Tetrahydrofuran (THF).

     

  • Connections: Fingertight connector

     

  • Flow rate: 0.01-100 ml/min

     

  • Material:
    Polyetheretherketone (PEEK) Polytetrafluoroethylene (PTFE)
    Ethylenechlorotrifluoro-ethylene (ECTFE) Polyvinylidenefluoride(PVDF)
    Polyethylene (PE) Titanium Titanium alloy (Ti alloy) Aluminium oxide (Al
    oxide) Ruby Zirconia Super high resistent stainless steel

     

  • Operating pressure max.: 100 bar (10 MPa) (1450 psi)

     

  • Operating Temperature: 4°C-40°C

     

  • Power consumption: Max. 400 VA with all extra components connected

     

  • Protection class: IP 20 (Housing)

     

  • Relative humidity: 20%-95% (non condensing)

     

  • Viscosity Max.: 5 cP

     

  • Voltage: 100-120/220-240 V AC

     

M-925 Mixer:
  • Flow rate: 0.001-100 ml/min
  • Material: Polyetheretherketone (PEEK), Polytetrafluoroethylene (PTFE)
  • Operating
    pressure max: 100 bar (10 MPa) at 0.1 to 100 ml/min (12 ml chambers)
    250 bar (25 MPa) at 0.001 to 100 ml/min (0.6, 2 and 5 ml chambers) 350
    bar (35 MPa) at 0.001 to 5 ml/min (90 μl and 0.2 ml chambers) (1450 psi
    at 0.1 to 100 ml/min (12 ml chambers) 3625 psi at 0.001 to 100 ml/min
    (0.6, 2 and 5 ml chambers) 5075 psi at 0.001 to 5 ml/min (90 μl and 0.2
    ml chambers))
  • Operating Temperature: 4°C-40°C
  • pH Stability Long Term: 1-13
  • pH Stability Short Term: 1–14
  • Power consumption: 1 VA
  • Power requirements: 12 to 40 V DC from the pump
  • Protection class: IP 43
  • Relative humidity: 20%-95% (non condensing)
  • Viscosity Max: 5 cP
INV-907 Motor Valve:
  • Chemical stability: For use with all aqueous solutions and organic solvents commonly employed in chromatography
  • Connections: UNF 10-32 2B fingertights for capillary tubing o.d. 1/16″
  • Flow Channel i.d.: 0.8 mm
  • Flow rate: < 100 ml/min
  • Material: Polyetheretherketone (PEEK)
  • Operating Temperature: 4°C-40°C
  • pH Stability Long Term: 1-13
  • Power consumption: 1.1 VA
  • Power requirements: 32 VDC from Pump P-901
  • Relative humidity: 20%-95%
  • Viscosity Max.: <5 cP
  • Wetted Materials:  Polyetheretherketone (PEEK)
  • Wetted Part: Yes
  • Number of Ports: 7
  • Operating Pressure Max: 250 bar (25 MPa) (3626 psi)
P-950 Sample Pump:
  • Flow Rate: Up to 50 ml
  • Maximum Pressure: 1 MPa
  • Operating Temperature: 4° to 40°C

CU-950 Control Unit:

  • Guidance:
    1.8 m output cable with 5.5mm outer and 2.5 inner diameter DC connector
    Important: For CU-950 with new type of connector (1.8m output cable with
    3-pin KPPX3P connector) order Power Supply CU-950 28-9387-74
  • Wetted Part: N
Our business hours are Monday through Friday from 8:00 AM to 4:00 PM Eastern Time.  If you have questions, offers, or any other concerns regarding a listing outside our normal business hours, please allow for us to respond on the following business day. In the event you have any issues with your purchase, please contact us first. Customer satisfaction is our highest priority.
Once sale is completed item will be shipped within 4-5 business days of purchase.
 
JB
Code: TM

 

Additional information

Weight 113.40 lbs
Model

Modified Item

MPN

Brand

Chromatography Technique

Intended Use/Discipline

260 Interstate Circle
Suite 2A
Frederick , MD 21704

Call us

301 803 9722
240 375 6333