Description
Agilent 1100 Series HPLC System
This item comes with a 30-day Repair or Replace warranty.
This HPLC system is in very good condition. All operational testing was completed successfully, and each unit functions as intended. We used in-house software for testing. Please Note: If you need computer and software, please contact us for information prior to ordering, otherwise will come only with what is shown.
G1313A – Autosampler
- Injection Range: 0.1 µL to 100 µL, with 0.1 µL increments
- Replicate Injections: 1 to 99 per vial
- Sample Capacity: 100 x 2 mL vials per tray
- Minimum Sample Volume: 1 µL from a 5 µL sample in a 100 µL micro vial
- Injection Cycle Time: Typically, 50 seconds
- Carryover: Typically, < 0.01% to < 0.05%
- Sample Viscosity Range: 0.2 cp to 50 cp
- Physical Dimensions: 200 mm (height) x 345 mm (width) x 435 mm (depth)
- Weight: 14.2 kg (31.3 lbs)
- Power Requirements: 100-120 or 220-240 VAC, ± 10%, 50 or 60 Hz
G1316A – Thermostatted Column Compartment (TCC)
- Temperature Range: 10°C below ambient to 80°C
- Temperature Stability: ± 0.15°C
- Temperature Accuracy: ± 0.8°C
- Column Capacity: Three 30 cm columns
- Warm-up/Cool Down Time: 5 minutes from ambient to 40°C
G1315B – Diode Array Detector (DAD)
- Detection Type: 1024-element photodiode array
- Light Source: Deuterium and tungsten lamps
- Wavelength Range: 190 – 950 nm
- Short Term Noise: ± 1 x 10^-5 AU at 254 and 750 nm
- Drift: 2 x 10^-3 AU/hr at 254 nm
- Linear Absorbance Range: > 2 AU (upper limit)
- Wavelength Accuracy: ± 1 nm
- Slit Width: Programmable slit widths of 1, 2, 4, 8, and 16 nm
- Flow Cells: Various options including standard, semi-micro, micro, high pressure, and nano flow cells
- Control and Data Evaluation: Agilent ChemStation for LC
- Analog Outputs: Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, two outputs
- Communications: Controller-area network (CAN), GPIB, RS-232C, APG Remote, LAN optional
- Safety and Maintenance: Extensive diagnostics, error detection, leak detection, and safe leak handling
- Power Requirements: 100 – 240 VAC, 50 or 60 Hz, 300 VA / 125 W / 427 BTU
G1312A – Binary Pump
- Weight: 34 lbs.
- Dimensions: 7 x 13.5 x 17 in.
- Electrical: 100 – 240 VAC, ± 10% ~ 50/60Hz ~ 220VA ~ 74W ~ 253BTU
- Ambient Operating Temp: 4–55 °C (41–131 °F)
- Ambient Non-Operating Temp: -40–70 °C (-4–158 °F)
- Maximum Operating Pressure: 600 bar (8700 psi)
- Flow Rate Range: 0.1 – 5 mL/min
- Flow Precision: < 0.07% RSD or < 0.02 min SD
- Flow Accuracy: ± 1% or 10 µL/min
G1379A – Micro Vacuum Degasser
- Channels: 4-channel vacuum container
- Membranes: 4 micro-structured membranes
- Vacuum Pump: Integrated
- Internal Volume: 480 µL per channel
- Maximum Flow Rate: 10 mL/min per channel
- Operating Temperature: 4–55 °C (41–131 °F)
- Power Consumption: 30 VA