Description
Millipore Mobius CellReady 200L Bioreactor Skid System
60 day limited warranty included~
-
Dell Precision T3500 Workstation Intel Xeon 3500 Software: TruBio™ powered by Emerson™ DeltaV TruLogic Controller.
-
Dell 2007FPb 20.1″ Flat Panel Monitor 1600 x 1200 @ 60 Hz (UXGA)
This system is in very good condition. Please see description for all components that will come with this purchase. Manufacture date: 2011.
- Working Volume: 40–200 liters
- Total Volume (including headspace): 250 liters
- Aspect Ratio (Height/Diameter): 2.0:1
- Bioreactor Diameter: 21.3 in.
- Maximum Operating Pressure: 17 mbar (0.25 psi)
- Operating Temperature Range: 4°C to 40°C
- Impeller Type: Pitched blade (4 blades at 13° pitch)
- Impeller Diameter: 18.28 cm (7.2 inches)
- Impeller Position: Bottom-mounted, 15° offset from center
- Motor Speed Range: 30–140 rpm
- Power Number: 4.0
- Internal Baffle: X-baffle design for enhanced mixing
- Vent Filter: Opticap® XL5 capsule with Aervent® 0.2 µm membrane
- Vent Filter Flow Rate: 80 LPM @ 0.25 psi
- Overlay Filter: Millipak® with 0.22 µm hydrophobic Durapore® membrane
- Microsparger Filter: Opticap® XL300 with 0.2 µm Millipore Express® SPG membrane
- Mobius® SensorReady Technology:
- Supports modular sensor assemblies
- Compatible with magnetic levitating bearing less pumps
- Non-invasive flow sensors
- Lynx® S2S sterile connectors for closed-loop sampling
Lauda T 1200 Variocool Circulation Chiller
- Cooling Capacity: Approximately 1.2 kW (1200 W)
- Temperature Range: -20°C to +40°C (standard); optional heaters can extend this up to +80°C
- Cooling Method: Air-cooled or water-cooled (model-dependent)
- Control Interface: Color TFT display
- Membrane keypad
- USB interface and alarm contact standard
- Connectivity: Optional interfaces for integration with external systems
- Design: Compact footprint; models up to 2 kW can fit under lab benches
- Electronic fill gauge and low-level alarm
- Tool-free removable front grid for easy cleaning
- Adjustable bypass and pressure gauge (on models VC 1200 and above)
Mettler Toledo IND331 Terminal
- Maximum Cable Length: 300 m (1000 ft.)
- Min/Max Load Resistance: 500 Ω maximum
- Resolution: 16-bit resolution – 65536 levels across entire range
- Error Indication: Two open collector error signals to indicate error conditions of the terminal. Labeled as Under and Over
Mobius® CellReady Full Size Controller Cart
- Houses and integrates the control hardware for Mobius CellReady bioreactors.
- Control Software: TruBio™ software powered by Emerson™ DeltaV
- Architecture: Open, modular, and scalable
- Compatibility: Designed for 200 L Mobius CellReady bioreactor systems.
- Mobius SensorReady™ Technology: Supports modular sensor and pump assemblies for real-time process monitoring and control.
- Bearing less Magnetic Drive Pumps
- Non-invasive Flow Sensors
- Lynx® S2S Sterile Connectors
- Dual or Quad RS (Remote Sensor) Assemblies
- Mobility: Mounted on lockable casters for easy movement
- Data Interfaces: USB, Ethernet, and analog/digital I/O for integration with SCADA or DCS systems
- Operating Temperature: 15–30°C
- Humidity: 20–80% non-condensing
TruLogic Controller 1002092-021
-
- Dimensions (DxWxH): 24 x 8 x 24 in.
- Rating NEMA 4X
- Operating Temperature 5 –40°C (41 – 104°F)
- Storage Temperature -25 –70°C (-15 –158°F)
- Relative Humidity 5– 95% (non-condensing)
- Power 110-240 VAC, 50/60 Hz
- Certification CE (EN-61326 and EN-61010) and UL-508
- Dell Precision T3500 Workstation Intel Xeon 3500 Software: TruBio™ powered by Emerson™ DeltaV TruLogic Controller.