Agilent 1200 Series HPLC System w/Computer and Software

$18,750.00

1 in stock

SKU: 010580 Category:

Description

Agilent 1200 Series HPLC System w/Computer and Software

This Agilent 1200 Series System is in very good condition. Retrieved from working lab; we’ve tested components. Will come with computer and ChemStation software.

G1337A – Micro Well-Plate Autosampler (µWPS) 

  • Injection Range: 0.01–8 µL (small loop) or 0.01–40 µL (extended loop) in 0.01 µL increments. 
  • Pressure Range: Up to 400 bar (5880 psi). 
  • Precision: Typically, <0.5% RSD (5–40 µL); <1% RSD (1–5 µL); <3% RSD (0.2–1 µL). 
  • Sample Capacity: 2× well-plates (MTP) + 10 × 2 mL vials (using standard tray). 
  • Carry-over: Typically, <0.05%. 
  • Injection Cycle Time: Typically, <30 seconds. 
  • Sample Viscosity Range: 0.2–5 cP. 
  • Dimensions (DxWxH): 17 x 13.5 x 8 in. 
  • Weight: 34.2 lbs.

G1330B – Autosampler Thermostat 

  • Temperature Range: 4°C to 40°C 
  • Technology: Peltier-controlled heat exchanger for cooling and heating 
  • Dimensions (H x W x D): Approximately 5.5 x 13.5 x 17 inches 
  • Weight: Approximately 46 lbs  
  • Voltage: 100-120 or 220-240 VAC 
  • Line Frequency: 50 or 60 Hz 
  • Power Consumption: Approximately 260 VA for the thermostat   

G1316A – TCC 

  • Weight: 25 lbs.   
  • Dimensions (HxWxD): 5.5 × 13.5 × 17 in.   
  • Line Voltage: 100 – 240 VAC, ± 10%   
  • Line Frequency: 50 or 60 Hz, ± 5%   
  • Power Consumption: 220 VA, 85 W / 290 BTU  
  • Ambient Operating Temp: 0–55°C (32–131°F)   
  • Ambient Non-Operating Temp: -40–70°C (-4–158°F)   
  • Humidity: < 95%, at 25–40 °C (77–104 °F) non-condensing   
  • Operating Altitude: Up to 2000 m (6500 ft)   
  • Non-Operating Altitude: Up to 4600 m (14950 ft) For storing the instrument   
  • Safety Standards: IEC, CSA, UL, EN   
  • Installation Category II, Pollution Degree 2

G1315B – Diode Array Detector (DAD) 

  • Detector Type: 1024 element photodiode array. 
  • Wavelength Range: 190 – 950 nm. 
  • Light Source: Deuterium (UV) and Tungsten (VIS) lamps. 
  • Short Term Noise (ASTM): ± 1 × 10-5 AU at 254 and 750 nm 
  • Drift: 2 × 10-3 AU/hr. at 254 nm  
  • Linear Absorbance Range: > 2 AU (upper limit) 
  • Wavelength Accuracy: ± 1 nm 
  • Slit Width: Programmable 1, 2, 4, 8, 16 nm. 
  • Data Rate: Up to 80 Hz. 
  • Flow Cell (Standard): 13 \(\mu\text{l}\) volume, 10 mm cell path length, 120 bar pressure maximum.

G1379B – 4-channel Micro Vacuum Degasser 

  • Maximum Flow Rate: 0 – 5 ml/min per channel. 
  • Number of Channels: 4. 
  • Internal Volume per Channel: Approximately 1 ml. 
  • Materials in Contact with Solvent: PTFE, FEP, and PEEK. 
  • pH Range: 1 – 14. 
  • Dimensions (WxDxH): 13.5 x 17 x 3.1 in. 
  • Weight: 16 lbs.  
  • Electrical Requirements: 100 – 240 VAC, ±10%, 50/60 Hz, ±5%.  
  • Power Consumption: 30 VA / 30 W / 102 BTU Maximum.  
  • Ambient Operating Temperature: 0 – 55 °C (32 – 131 °F). 
  • Operating Altitude: Up to 2000 m (6562 ft). 
  • Humidity: < 95% non-condensing.

G1376A – Capillary Pump 

    • Flow Range: 0.01 – 20 µL/min (standard), 0.01 – 100 µL/min (with extended flow range kit). 
    • Pressure Range: Up to 400 bar (5880 psi). 
    • Flow Accuracy: < 0.7% RSD or 0.03% SD (at 10-50 µl/min 
    • Delay Volume: Low delay volume for fast gradients. 
    • pH Range: 1.0 – 8.5 
    • Dimensions (DxWxH): 17.2 x 13.5 x 7.1 in. 
    • Weight: 38 lbs. 

Ebay