Promega 9311.011/E9031 GloMax Multi+ Microplate Multimode Reader
Item is in good condition. Item was tested for Luminescence, Fluorescence (Intensity), Visible Absorbance, UV-Visible Absorbance. This item also has the E9031 heating and shaking features. Plates have not recently been calibrated.
Detection Modes: Luminescence, Fluorescence, Visible Absorbance and UV-Visible Absorbance
- Sample Format: 6-, 12-, 24-, 48-, 96- and 384-well microplates
- Heater Maximum: 45°C ± 0.75°C
- Heating Interval: 0.1–120 minutes
- Integrated PC Operating System: Windows® CE 6
- User Interface: 6.6” VGA color TFT LCD display with touch screen
- Data Output: PC-compatible ascii file format
- Linear and Orbital Shaking Pattern
- Shaking Intensity: Low, Medium, High and Variable, (150, 300, 500rpm, and a combination of the three, respectively)
- Shaking Interval: 0.1–120 minutes
- Dimensions: 19″x18″x14″
- Weight: 43 lb
- Power Requirements: AC Power Adapter: Input: 100–240VAC, 50/60Hz Output: 24VDC, 6.3A, 150W Max.
- Instrument: Input: 24VDC, 5.5A Max
Luminescence Module Specifications
- Detector: Top-reading, head-on photon counting photomultiplier tube (PMT)
- Wavelength Range: 350–650nm
- Detection Limit: 3 × 10–21 moles of luciferase
- Linear Dynamic Range: >8 logs
- Crosstalk: 1 million RLU using 96-well plates White Plates: 10–5 (for the Corning Costar Cat.# 3789) 10–4 (for the Greiner Bio-One LumiTrac 200) Black Plates: <10–6
Absorbance Module Specifications
- Light Source: LED
- Detector: Top-reading, large-area photodiode
- Spectral Range: 450–750nm
- Filters Wheel Capacity: Holds up to six filters: four installed filters and two empty filter holders for user configuration
- Wavelengths Included: 450, 560, 600, 750nm
- Photometric Measuring Range: 0–5.0 OD
- Linear Dynamic Range: 0–4.0 OD
- OD Accuracy: 0.01 OD ± 3%
- OD Precision: 0.01 OD ± 1%
- Stray Light: 0.002% at 560nm in clear bottom black-wall plates
UV-Visible Absorbance Module Specifications
- Light Source: Xenon Flash Lamp
- Detector: Top-reading photodiode
- Spectral Range: 200–1,100nm
- Filters Wheel Capacity: Holds up to six filters: four installed filters and two customizable filter paddles with 260nm and 280nm preinstalled
- Wavelengths Included: 260, 280, 450, 560, 600, 750nm
- Photometric Measuring Range: 0–5.0 OD
- Linear Dynamic Range: 0–3.0 OD, OD Accuracy: 0.05 OD ± 3%, OD Precision: 0.002 OD ± 3%, Stray Light: 0.001% at 560nm in clear-bottom, black-walled plates
Fluorescence Module Specifications
- Light Source: Wavelength matched LED
- Detector: PiN-photodiode
- Read Position: top reading
- Wavelength Selection: snap-in optical kits
- Wavelengths Included: UV (Ex: 365nm, Em: 410–460nm) Blue (Ex: 490nm, Em: 510–570nm), Green (Ex: 525nm, Em: 580–640nm), Red (Ex: 625nm, Em: 660–720nm)
- Detection Limit: 0.5fmol/200µl or 1ppt of fluorescein in 96-well microplate
- Linear Dynamic Range: 6 logs (assay-dependent)