Description
SciLog SciPure 200 TFF with Weigh Station and Terminal
This item comes with a 30-day Repair or Replace warranty.
SciPure 200 Tangential Flow Filtration System 900-200-8
- Applications: Concentration, diafiltration clarification / sterilization, media and buffer preparation, fluid transfer, mixing, chromatography and column loading, harvest, desalting, viral filtration
- Inlet Flow Range: 1 – 20 L/min
- Maximum Pressure: 30 psi
- Hold-Up Volume: Less than 500 mL
- Recommended Filter Area: 0.1 – 2.5 m² (upgradable to 5 m²)
- Touchscreen Display: Full color graphics and menu-driven operation
- Other Utilities: Compressed Air, Nitrogen @ 35 PSI
- Enclosure & Rating: 304 Stainless Steel framework, Mobile platform with pharmaceutical grade casters, NEMA 4X, IP56
- I/O Ports: Standard connections for 3 SciPres pressure sensors, 2 SciCon conductivity sensor, retentate quantification and permeate flow rate / quantification. Additional I/O and digital communications available.
- Operational Mode: Endpoint concentration and diafiltration modes, manual mode
- International Quality Standards: RoHS, CE, ETL, CSA and UL
- Plant Communications: Rockwell Ethernet IP
- Power Requirements: 208VAC 1/3 Ph 20A, 60
- Physical Dimensions (LxWxH): 48.5 x 30 x 48 in.
SciLog Weight Station 900-WSTN
- Maximum Load: 50Kg (ie. 35L Bioprocess Container)
- Resolution: 10g
- Dimensions: 29 x 29 in.
- Materials of Construction 304 Stainless Steel, Pharmaceutical Grade Casters
- Power: 120 VAC ~ 1 Ph ~ 60Hz ~ < 2 AMP
IND331 Harsh Terminal
- Application: Process weighing
- Housing: Stainless steel enclosure
- Display: OLED, 12 mm high
- Keyboard: Zero, tare, clear, print
- Resolution: 100,000 divisions
- Dimensions: 8.7 x 5.2 x 7 in.
- Degree of Protection: IP66
- Power Supply: Universal AC or external DC