Agilent 1100 Series HPLC System
This item comes with a 30-day Repair or Replace warranty.
This HPLC system is in very good condition. All operational testing was completed successfully, and each unit functions as intended. We used in-house software for testing. Please Note: If you need computer and software, please contact us for information prior to ordering, otherwise will come only with what is shown.
G1329A – Autosampler
- Injection Range: 0.1 to 100 µL
- Injection Precision: < 0.5% RSD
- Sample Capacity: 100 × 2 mL vials or 2 × 96 well plates
- Carryover: < 0.05%
- Injection Cycle Time: 15 seconds
- Temperature Control: Optional, 4-40°C
- Dimensions (DxWxH) : 13.6 x 17.1 x 7.1 in
- Weight: 24 lbs.
G1330B – Autosampler Thermostat
- Operating Temperature Range: 4 to 40°C (39 to 104°F)
- Operating Humidity Range: < 95%, non-condensing
- Maximum Power Consumption: 260 VA (210 W)
- BTU: 717
G1316A – TCC
- Temperature Range: 10°C below ambient to 80°C
- Temperature Stability: ± 0.15°C
- Temperature Accuracy: ± 0.8°C
- Column Capacity: Three 30 cm columns
- Warm-up/Cool Down Time: 5 minutes from ambient to 40°C
G1315B – DAD
- Detection Type: 1024-element photodiode array
- Light Source: Deuterium and tungsten lamps
- Wavelength Range: 190 – 950 nm
- Short Term Noise: ± 1 x 10^-5 AU at 254 and 750 nm
- Drift: 2 x 10^-3 AU/hr at 254 nm
- Linear Absorbance Range: > 2 AU (upper limit)
- Wavelength Accuracy: ± 1 nm
- Slit Width: Programmable slit widths of 1, 2, 4, 8, and 16 nm
- Flow Cells: Various options including standard, semi-micro, micro, high pressure, and nano flow cells
- Control and Data Evaluation: Agilent ChemStation for LC
- Analog Outputs: Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, two outputs
- Communications: Controller-area network (CAN), GPIB, RS-232C, APG Remote, LAN optional
- Safety and Maintenance: Extensive diagnostics, error detection, leak detection, and safe leak handling
- Power Requirements: 100 – 240 VAC, 50 or 60 Hz, 300 VA / 125 W / 427 BTU
G1311A – Quaternary Pump
- Power: 100 – 240 VAC, ± 10% ~ 50/60Hz ~ 180VA ~ 55W ~ 188BTU
- Ambient Operating Temp: 4–55 °C (41–131 °F)
- Ambient Non-Operating Temp: -40–70 °C (-4–158 °F)
- Humidity: < 95%, at 25–40 °C (77–104 °F), non-condensing
- Operating Altitude: Up to 2000 m (6500 ft)
- Hydraulic System: Dual plunger in series pump with proprietary servo-controlled variable stroke drive, floating plungers, and active inlet valve
- Settable Flow Range: 0.001 – 10 ml/min, in 0.001 ml/min increments
- Flow Precision: ≤ 0.07% RSD, or ≤ 0.02 min SD, whichever is greater, based on retention time at constant room temperature
- Flow Accuracy: ± 1% or 10 μl/min, whichever is greater
- Pressure Operating Range: 0 – 40 MPa (0 – 400 bar, 0 – 5880 psi) up to 5 ml/min; 0 – 20 MPa (0 – 200 bar, 0 – 2950 psi) up to 10 ml/min
- Pressure Pulsation: < 2 % amplitude (typically < 1 %), at 1 ml/min isopropanol, at all pressures > 1 MPa (10 bar)
- Compressibility Compensation: User-selectable, based on mobile phase compressibility
- Recommended pH Range: 1.0 – 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel
- Gradient Formation: Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve
- Delay Volume: 800 – 1100 μl, dependent on back pressure
- Composition Range: 0 – 95 % or 5 – 100 %, user selectable
- Composition Precision: < 0.2 % RSD, at 0.2 and 1 ml/min
- Communications: Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional
G1322A – Vacuum Degasser
- Channels: 4-channel vacuum container
- Components: Includes 4 tubular plastic membranes and a vacuum pump
- Function: The vacuum pump generates a partial vacuum in the vacuum container, which is measured by a pressure sensor