Description
Agilent 1200 Series HPLC System w/Software and Computer System
This HPLC system is is very good condition. We’ve tested all functionalities of the components, and software. 60 Day Warranty.
G1367B – High Performance Autosampler
- Injection Range: 0.1–100 µL. Up to 1500 µL with multiple draw and hardware modification.
- Precision: Typically, <0.25% Relative Standard Deviation (RSD) from 5–100 µL.
- Pressure Range: Up to 400 bar (5880 psi).
- Sample Viscosity: 0.2 – 5 cp.
- Sample Capacity: Holds 2× well plates (MTP), plus additional vials and Eppendorf tubes in various configurations.
- Weight: 15.5 kg (34.2 lbs).
- Dimensions (HxWxD): 8 × 13.5 × 17 in.
- Line Voltage: 100–240 VAC, ±10% wide-ranging capability.
- Ambient Operating Temperature: 4 to 55 °C (41 to 131 °F).
G1330B – Autosampler Thermostat
- Temperature Range: 4°C to 40°C
- Technology: Peltier-controlled heat exchanger for cooling and heating
- Dimensions (H x W x D): Approximately 5.5 x 13.5 x 17 inches (14 x 34.5 x 43.5 cm)
- Weight: Approximately 46 lbs (20.7 kg)
- Voltage: 100-120 or 220-240 VAC
- Line Frequency: 50 or 60 Hz
- Power Consumption: Approximately 260 VA for the thermostat
G1316A – TCC
- Weight: 25 lbs.
- Dimensions (HxWxD): 5.5 × 13.5 × 17 in.
- Line Voltage: 100 – 240 VAC, ± 10%
- Line Frequency: 50 or 60 Hz, ± 5%
- Power Consumption: 220 VA, 85 W / 290 BTU
- Ambient Operating Temp: 0–55°C (32–131°F)
- Ambient Non-Operating Temp: -40–70°C (-4–158°F)
- Humidity: < 95%, at 25–40 °C (77–104 °F) non-condensing
- Operating Altitude: Up to 2000 m (6500 ft)
- Non-Operating Altitude: Up to 4600 m (14950 ft) For storing the instrument
- Safety Standards: IEC, CSA, UL, EN
- Installation Category II, Pollution Degree 2
G1314C – Variable Wavelength Detector
- Wavelength Range: 190-600 nm
- Wavelength Accuracy: ±1 nm (self-calibrated with deuterium lines and verified with a holmium oxide filter)
- Wavelength Bandwidth: 6.5 nm typical
- Data Rate: Up to 55 Hz
- Noise (ASTM, 254 nm): < ± 0.5 x 10⁻⁵ AU
- Drift (254 nm): < 3 x 10⁻⁴ AU/hr.
- Linearity: > 2.0 AU upper limit
- Detection Type: Double-beam photometer
- Light Source: Deuterium lamp
- Automatic Calibration: Holmium oxide filter for wavelength verification
- Standard: 14 µL volume, 10 mm path length, 40 bar max pressure
G1379B – 4-channel micro vacuum degasser
- Maximum Flow Rate: 0 – 5 ml/min per channel.
- Number of Channels: 4.
- Internal Volume per Channel: Approximately 1 ml.
- Materials in Contact with Solvent: PTFE, FEP, and PEEK.
- pH Range: 1 – 14.
- Dimensions (WxDxH): 13.5 x 17 x 3.1 in.
- Weight: 16 lbs.
- Electrical Requirements: 100 – 240 VAC, ±10%, 50/60 Hz, ±5%.
- Power Consumption: 30 VA / 30 W / 102 BTU Maximum.
- Ambient Operating Temperature: 0 – 55 °C (32 – 131 °F).
- Operating Altitude: Up to 2000 m (6562 ft).
- Humidity: < 95% non-condensing.
G1312A – Binary Pump
-
- Flow Rate: 0.001 to 5.0 mL/min, in 0.001 mL/min increments.
- Flow Precision: ≤0.07% RSD (typical).
- Flow Accuracy: ±1% or 10 µL/min (whichever is greater).
- Pressure Limit: 0 to 400 bar (0 to 5880 psi) at 5 mL/min.
- Gradient Capability: High-pressure binary mixing for precise gradients.
- Hydraulic Design: Two-channel, dual-piston in-series design with floating pistons and active inlet valves.
- Mobile Phase Compatibility: Recommended pH range of 1.0 to 12.5.
- Dwell Volume: Low, with a range of 180 to 480 µL without a mixer.
- Internal Components: Includes a purge valve with a PTFE frit, a low volume mixing chamber, a damping unit, and an optional mixer.
- Electrical: Auto-ranging 100–240 VAC, 50/60 Hz wide-ranging capability.
- Weight: 34 lbs.
- Dimensions (HxWxD): 7 x 13.5 x 17 in.