Description
Kuhner SB10-X Single-Use Bioreactor w/ES-X Lab Shaker & Control Unit
This shaken bioreactor system is from a working lab, in working condition. Will come with all components pictured and connection cables. Embedded software is included.
The SB10-X is an orbital shaken benchtop bioreactor system for the cultivation of human, mammalian, and plant cells in a single-use bag. It consists of six modules (excluding the bag) and has a working volume from 4 liters up to 12 liters.
The SB10-X can be used in research, process development, or as a pilot scale bioreactor.
Shaken Bioreactor
- Working Volume: 4 L – 12 L (1.5 – 4.5 L)
- Total Bag Volume: approx. 24 L (approx. 6 L) (not included)
- Weight (bioreactor incl. shaker): approx. 166 lbs. (without liquid)
- Required Footprint (excl. control unit): 26.3 x 25.4 in.
Shaker Module ES-X
- Motor: Modified ES-X benchtop shaker
- Shaking Diameter: 50 mm (orbital motion) shaking diameter is adjustable (12.5 mm, 25 mm, 50 mm)
- Shaking Speed: 20 rpm – 140 rpm (when used with SB10-X)
- Loading Max: 55 lbs.
- Setting: Digital 1 rpm
- Accuracy, absolute: ± 0.5 rpm
- Active Brake: adjustable
- Ambient Temperature: 5°C – 40°C
- Casing: stainless steel
Mains Connection: 110 – 120 V / 50-60 Hz ~ 102024 95 – 105 V / 50-60 Hz
Temperature Control
- Heating resistance heater (100 W), integrated in tray module
- Temperature Max: 40 °C
- Temperature Min: approx. ambient temperature (no cooling)
- Temperature Accuracy: ± 0.25 °C (at 37 °C)
- Temperature Distribution: 8 mm thick aluminum plate provides homogenous temperature distribution of resistance heater
pH and DO control Integrated in tray module
- pH Measurement Principle: Optical chemosensor
- Range: pH 5.5 – pH 8.5
- Accuracy (chemosensor): ± pH 0.05 at pH 7 with one point calibration ± pH 0.10 at pH 7 with pre-calibration
- Drift: < pH 0.005 per day
- Temperature Range: up to 50 °C
- DO Measurement Principle: optical chemosensor
- Range: 0 % – 100 % DO
- Accuracy (chemosensor): ± 0.1 % O2 at 20.9 % O2 Accuracy (system) ± 10 % DO Drift < 0.015 % O2 per day
- Temperature Range: up to 50 °C
Filter Heater
- Capacity: 1 exhaust filter
- Casing: Polycarbonate
- Heating resistance heater (6 W)
- Max. Temperature: approx. 40 °C at 23 °C ambient temperature
- Control orange LED light indicates that filter heater is active and working
Control Unit
- Weight: approx. 60 lbs.
- Required Footprint: 27.4 x 12.9
- Casing: Stainless steel 1.4301
- Interface: 2 x CAN-Bus Optional: EMI-60, NET-60
Touch Screen
- Display: 12.1’’ screen (resistive) front is splash proof and dust protected (IP 64)
- Resolution: 1024 x 768
- Cooling: passive with no fan
- Main Memory: 2 GB (1.6 GB usable)
- Interfaces: 4 x USB, 2 x RS 232, 2 x LAN, 1 x VGA
Software – Kuhner Insight Application Software
- Monitoring: pH & DO, shaking speed, temperature, gas flow and mixture, pressure Calibration: Pre-calibrated sensors in bag
- Recalibration: One point recalibration is possible
- Applicable: Win 7 & 10
- GMP: 21 CFR Part 11 compliant
Pumps 3 Peristaltic Pumps
-
- Type: Watson-Marlow 313 VBM/D
- Rollers: 3
- Power: 70 W
- Maximum Speed: 100 rpm
- Tube: 1.6 mm wall thickness
- Delivery Pressure: Max. 2 bar